Your basket is empty
Home > Environmental > Nutrient Monitoring > EXO Sensors > EXO Wiped Conductivity and Temp. Sensor

EXO Wiped Conductivity and Temp. Sensor
Enlarge

EXO Wiped Conductivity and Temp. Sensor

PART NO: X-599827

Designed and engineered to be compatible with EXO2’s Central Wiper, the EXO Wiped Conductivity and Temperature Sensor is perfect for long-term monitoring applications in fouling environments.

Biofilms and algal growths can sometimes compromise your data quality, clogging up conductivity cells and coating sensor optics. Not with this sensor! Most conductivity cells cannot be fully cleaned by mechanical anti-fouling wipers, but the EXO Wiped (C/T) probe was purposely-built to protect your conductivity data.

Main Advantages of the EXO Wiped Conductivity and Temperature Sensor:

  • Save money by reducing the number of site visits each year. Pays for itself in under a year depending on application
  • Less cleaning & maintenance of conductivity sensors and anti-fouling screens (specially with removing the build-up of hard growth)
  • Improve the reliability of your conductivity sensor, now with non-metallic electrodes
  • Improve the conductivity data by avoiding stagnant readings and reducing the impact of micro-environments
  • Less post-processing and manually adjusting data for fouling-related sensor drift
  • Prevent common types of fouling from impacting your valuable data, including; particulates, algae, barnacles, and trapped gasses
  • Please know that the EXO Wiped Conductivity and Temperature Sensor requires an EXO2 sonde with central wiper installed to function.

    Technical Specifications

    Anti Fouling

    Yes, with EXO2 Central Wiper

    Depth / Pressure Rating / Limit

    250 meters

    Desktop Software Compatible

    Kor Software

    Equipment used with

    EXO2 with a Central Wiper installed, and Spacing Kit (599831)

    Monitoring

    Yes

    Operating Temperature

    -5 to 50°C

    Parameters Measured

    Conductivity
    Range: 0 to 100 mS/cm
    Accuracy: ±1% of the reading or 0.002 mS/cm, whichever is greater
    Resolution: 0.0001 to 0.01 mS/cm (range dependent)


    Specific Conductance
    Range: 0 to 100 mS/cm
    Accuracy: ±1% of the reading or 0.002 mS/cm, whichever is greater
    Resolution: 0.0001 to 0.01 mS/cm (range dependent)


    Salinity
    Range: 0 to 70 ppt
    Accuracy: ±2% of the reading or 0.2 ppt, whichever is greater
    Resolution: 0.01 ppt


    Temperature
    Range: -5 to 50°C
    Accuracy: ±0.2°C
    Resolution: 0.001°C 
    Response Time: T95<30 seconds

    Sampling

    No

    Smart Sensors / Ports

    Yes

    Storage Temperature

    -20 to 80°C

    User Replaceable

    Yes

    Warranty

    2 years

    Waterproof

    Yes

     

    Download Spec Sheet

    Catalogue NumberENVEMS0000232
    Manufacturer P/n599827
    Your basket is empty