Your basket is empty
BROWSE OUR PRODUCT CATALOGUE

Gas analyzers MAT1100/MAT1200

Gas analyzers MAT1000 series are used for measuring oxygen concentrations (O2) or optional carbon dioxide concentrations (CO2).
Applies to all models of MAT1000 series:

  • Easy and intuitive operation via touchscreen, even by untrained personnel
  • Create your own method
  • Available method parameters: scales, measurement procedures, limit values, comment and much more
  • Selectable measurement methods: single or continuous
  • Several single measurement procedures: manual, automatic, min/max detection
  • Shortcut keys for most important functions
  • Short measurement period, high accuracy even with small sample volume
  • Validation possible
  • Compliance with global standards such as GMP/GLP and 21 CFR Part 11
  • Documentation of all measured values of the last 999 measurements
  • Integrated interfaces, flexible data export (USB, Ethernet, RS-232)
  • Compact design in robust cast aluminum housing
  • Built-in air pressure sensor
  • Integrated flow rate monitoring

Gas analysers for controlling protective gas packaging. The MAT1100 equipped with an electrochemical oxygen cell measures the oxygen concentration within the range of 0.5 and 35 vol% and with an accuracy of ±0.2 vol%. The MAT1200 additionally features a non-dispersive infrared sensor for measuring carbon dioxide within the range of 0 and 50 vol% and with a high accuracy of ±0.5 vol%. Both devices are suitable for the random sampling as well as the permanent monitoring of the gas mixture used on packaging machines. Their solid metal housing with an interchangeable protective cover offers protection against impact and dirt from the outside.

Measurement of O2 and CO2 e.g. in modified atmosphere packaging, for work with modified atmosphere conditions.
  • Sporadic O2 sample checks
  • Quick Operation
  • Low-drift electrochemical oxygen cell (EC)
  • Low-drift carbon non-dispersive infrared sensor (NDIR), model MAT1200
  • Combined O2 and CO2 measurement, model MAT1200
  • Environmental compensated CO2 measurement, model MAT1200

Gas analyzers MAT1100/MAT1200
Specifications
O2 measuring range:
Accuracy:
Resolution:
0.5 to 35 Vol.-%
±0.2 Vol.-%
0.1 Vol.-%
CO2 measuring range (only MAT1200):
Accuracy:
Resolution:
0 to 50 Vol.-%
±0.5 Vol.-%
0.11 Vol.-%
Average measurment period:
Heating time MAT 1100 / MAT1200:
7 to 10 sec
<10 sec / 5 min
Interface:RS-232, USB, Ethernet
Dimensions (W x D x H):200 x 350 x 150 mm
Weight:4.6 kg
Power consumption:60 W
Power supply:100-250 V, 50/60 Hz

Gas analyzers MAT1100/MAT1200



A. Kruss Optronic Gas Analyzer MAT1100
PART NO.4AJ-6269834
77600||76949||78065
Type||Description||PK
MAT1100||with oxygen cell||1
A Kruss Optronic
MAT 1100
MAT1100

Gas analyzers

Sample bags, gaz analyzers

0
3
0
1
Out of Stock
A. Kruss Optronic Gas Analyzer MAT1200
PART NO.4AJ-6269835
77600||76949||78065
Type||Description||PK
MAT1200||with oxygen cell and NDIR carbon dioxide sensor||1
A Kruss Optronic
MAT 1200
MAT1200

Gas analyzers

Sample bags, gaz analyzers

0
3
1
1
Out of Stock

Copyright 2013 Lab Unlimited Sitemap | Terms & Conditions | ! Product images are for illustrative purposes only !